Fill the calibration pipe with fuel.
Immerse the Omnicomm LLS-HD sensor into fuel so that the probe is completely covered with fuel.
Wait until the green indicator lights up “The level is stabilized”. In the “Settings” tab in the “Full/Empty calibration” section press the “Full” button, and a value corresponding to the full tank will be recorded.
Remove the Omnicomm LLS-HD sensor from the tank and let the fuel drip from the probe. Please wait 1 minute at least. In the “Full/Empty calibration” section press the “Empty” button, and a value corresponding to the empty tank will be recorded.
Press the “Save to sensor” button.